Plasma Sterilizer

Low Temperature Plasma Sterilizer is user-friendly, as the main sterilizing agent is very safe, remaining nontoxic residue (Water and Oxygen.

STERİLMED 'PL series Plasma Sterilizers are offered with single door options in free standing configuration as a fully automatic plasma sterilizer. To assure fast & efficient sterilization, pre-set programs for different kinds of equipment with different specifications are set.

 

Item / Description

SM PL80

SM PL100

SM PL125

SM PL150

 

Sterilizing Agent

Hydrogen Peroxide

Hydrogen Peroxide

Hydrogen Peroxide

Hydrogen Peroxide

25 Cycles / Bottle

20 Cycles / Bottle

15 Cycles / Bottle

10 Cycles / Bottle

 

Total Cycle Time

Quick

25+5 Min

30+5 Min

40+5 Min

50+5 Min

Standard

35+5 Min

40+5 Min

50+5 Min

60+5 Min

Special

45+5 Min

50+5 Min

60+5 Min

70+5 Min

Cycle Temperature

50±5

50±5

50±5

50±5

SAL (Sterility Assurance Level)

10-6

10-6

10-6

10-6

 

Dimensions

Overall

680mm(W) x 1600mm(H) x 930mm(D)

680mm(W) x 1600mm(H) x 980mm(D)

680mm(W) x 1600mm(H) x 1030mm(D)

680mm(W) x 1600mm(H) x 1030mm(D)

Chamber

400mm(W) x 400mm(H) x 520mm(D)

400mm(W) x 400mm(H) x 640mm(D)

400mm(W) x 400mm(H) x 800mm(D)

400mm(W) x 400mm(H) x 940mm(D)

Volume

Total : 80 Liter

Total : 100 Liter

Total : 125 Liter

Total : 150 Liter

Weight

280kg

300kg

330kg

350kg

Lumen Claims

Rigid Lumen & Flexible, Lumen, Endoscopes

 

Chamber

Type

Rectangular

Material

Stainless Steel(SUS)

Control

Microprocessor & Windows CE Embedded

Cycle Information

Screen, Printer, USB, 100/10Mbps Ethernet(Option)

PM Cycle

Automatic Alarm & Setup

Electrical

110V/220V, 50/60Hz, 1Phase, 3000W

Installation Requirements

Front, Rear : 100cm

Left Side, Right Side: 10cm

Placement : Built-in Wheels Provide Mobility

Room Conditions

5 ~ 40 , 0 ~ 95%RH(Non-Condensing)

Printer

Built-in Thermal Printer(60Ø Roll Paper), Cycle Parameters (Temp, Pressure, Time, Daily&Total Cycle, Etc), Alarm & Error Display

Others

Emergency Stop(Front), Operator ID Login, Self-Diagnosis, Multi-Language, RFID Coding System

 

APPLICATION

STERİLMED PL series plasma sterilizers are used in cold and fast sterilization in hospitals, laboratories and dental clinics for all heat and moisture sensitive equipment which are resistant to H2O2.

COMPATIBILITY

Metal instruments and medical devices made of  aluminum,  brass  stainless  steel or titanium; plastic 1 medical devices made  of  Teflon  (PTFE  or  similar),  PE,  PU,  PVS,  silicone,  polystyrene,  PP,  PMMA, ULTEM,  Radel,  PEEK,  polycarbonate,  lexan,  marcalone,  LCP,  kraton,  EVA,  nylon,  polyamide,  delrin  or POM;  glass  materials.  Operator  is  responsible  for  sterilization  of  proper  materials  in  PLASMA  120 Series.  Sterilization  load  guides  provided  by  medical  device manufacturers  and  related institutes must be referred by operator before using device.

OVERALL

Device is designed without any PVC Lexan printed labels or stainless steel on the panel in order to reduce the risk of infection. All outer surfaces are from shock resistant glass surface (black or white color on request). Brand marking is done by logo placed behind tempered glass panel. Also manual controls are omitted from front panel design as all controls are established by a 12.1” Color TFT  touch screen. Firm fixing is done with suspension legs, which also enables leveling in non-flat surfaces. Movement ability is provided with wheel already installed.

TECHNICAL FEATURES

Integrated RF plasma system provides cold, fast and residue free sterilization. Plasma system is activated during cycle and turns H2O2 into O2 and H2O for a environmental friendly process. Weight reduced aluminum chamber coated with a specially formulated PTFE polymer.

Non-sticking and scratch resistant coating provides clean and brand new looking even after intense use. Each shelf inside chamber is able to carry 30 kgs of load. It is possible to open chamber door with a foot switch (sensor). Sterilization process is completely validatable. RFID system is used for operator identification and process safety. RFID tag cards are used by operators for process traceability.

  • Fast running cycle increases turnover rate the delicate and state-of-the art medical equipments in the hospital.
  • This rapid turnover rate lightens the hospitals’ financial burden, as they do not need to equip a number of redundant medical devices.
  • The by-products after sterilization, water and oxygen, contribute Green Environment as well as guarantee user-safety, substituting for other sterilizations which use harmful materials.
  • Low Temperature Plasma Sterilization is representative of all other kinds of Low Temperature Sterilization such as Ethylene Oxide or Formaldehyde.
  • The temperature keeps lower than 6oC and cycle time is under 1 hour, which prevents heat and moisture damages to sophisticated medical instruments, and prolongs the life expectancies of them.